Key parameters and specifications – Atec Agilent-4339B_4349B User Manual
Page 2
![background image](/manuals/318008/2/background.png)
2
Open:
Compensates leakage current
and capacitance in test fixtures
and cables.
Contact check:
Verifies reliability of
capacitor test connections.
Save/Recall:
Stores and retrieves up to
10 measurement states.
Comparator:
Selects values for high,
in, and low testing.
Measurement parameter:
Selects the right parameter
for the measurement.
Display:
LCD with back-light. Displays
measurement values, instrument
states, and comparator results.
Test sequence program:
Controls a series of resistance
measurements.
(Charge-measure-discharge)
Agilent 4349B 4-channel high resistance meter
Key parameters and specifications
Agilent 4339B
Agilent 4349B
Test channels
1
2ch, Option 4349B-001
4ch, Option 4349B-700
Test voltage(Vdc)
0.1 to 1000
Requires external power source
1
Measurement parameters
R, I, pv, ps
R, I
Measurement range
((Ω
Ω))
10
3
to 1.6x10
16
10
3
to 10
15
Basic accuracy
0.6% 2%
Display resolution
3 / 4 / 5 digits
3 / 4 / 5 digits
Measurement time
10 ms/30 ms/390 ms
9.5 ms/28 ms/98 ms/397 ms
1. External power source required for resistance measurements. Recommendation for external power source for
measurement for 1 G
Ω sample at 100 Vdc with accuracy = ±10%:
Ripple:
≤ 1 mVrms (50/60 Hz)
Wideband noise:
≤ 5 µVrms/Hz (50 Hz)
Switching noise:
≤ 50 mVrms (100 kHz)