Specifications
Measurement Accuracy

Agilent 4339B test conditions*:
1. Warm up time:

≥ 30 minutes

2. Ambient temperature: 23 °C ± 5 °C
3. Test cable length:

≤ 1.5 meters

4. Open error correction performed
5. Long measurement time setting
6. Contact check: off

Accuracy parameters:
Rm: Measured resistance value in W
Im: Measured current value in
amperes
Vs: Source voltage in volts
Vo: 0.1 V@Vs

≤ 200 V, 0.5 V@Vs > 200 V

Accuracy example:
To determine the accuracy of a meas-
urement use Figure 1, “Conversion
diagram”.
For example: determine the accuracy
of a 5 M

Ω (= 5 x 10

6

Ω) measurement

at 50 Vdc.
Rm = 5 x 10

6

Vs = 50 V

The intersection of Rm running paral-
lel to the 10

6

Ω diagonal line intersects

the vertical Vs line at the second row
from the bottom of the diagram. Mov-
ing horizontally across to Table 1, the
following equation is found:

100 Vo + (2.5 x 10

-7

x Rm)

0.5 +

Vs

Entering the values for Rm, Vo, and
Vs yields an accuracy of ±0.725%.

Figure1. Conversion diagram

Table 1. Agilent 4339B measurement accuracy (±% of reading)