Measurement parameters:
Fr, Fs, Fa, FL, CI, Q, C0, C1, L1, R1,
R0, G0, spurious, DLD, EM, filter
bandwidth
(Note that Fr and Fs represent the
frequencies at which the phase is 0
degree and the maximum conduc-
tance is attained, respectively. You
can also set up the HP E4916A to
search for Fr at which |Z| is minimal.)
Output
Frequency range: 1 MHz to 180 MHz/
1 mHz resolution
Frequency accuracy: 2 ppm (can be
improved up to 0.2 ppm through
external reference input)
Inputs (typical)
Measurement range: 1 MHz to
180 MHz/1 mHz resolution
Impedance: 5 to 125
Ω
/10 m
Ω
resolution
Return loss: 20 dB
Max. input level: 0 dBm
Damage level: 25 Vdc, 20 dBm
Average noise level: -130 dBm/Hz
Cross talk: -100 dB
Dynamic accuracy:
Input level
Level
Phase
accuracy
accuracy
0 to -20 dBm
0.5 dB
5°
-20 to -30 dBm
0.4 dB
3°
-30 to -40 dBm
0.1 dB
0.6°
-40 to -50 dBm
0.05 dB
0.3°
-50 to -60 dBm
0.05 dB
0.3°
-60 to -70 dBm
0.1 dB
0.6°
-70 to -80 dBm
0.5 dB
5°
-80 to -90 dBm
1.0 dB
10°
General Specifications
Power supply: 90 to 132 V or 198 to
264V, 47 to 63 Hz, 100 VA max.
Operating temperature/humidity:
0 to 55
°
C/15 to 95%
Outer dimensions: 320W
×
100H
×
450D mm
Weight: 5.3 kg
Specifications
Others
Interface: HP-IB interface, handler
interface
External reference input: 1/2/5/10
MHz,
±
10 ppm
Equipment supplied with the HP
E4916A:
2 BNC cables (120 cm), 1 operation
manual, 1 user's guide, 4 sample
program disks (VEE run-only
versions included), 1 power cable
Measurement Accuracy
Resonator testing:
Fr: 2 ppm (SPC: Supplemental
Performance Characteristics)
CI: 5% (SPC)
LCR measurements:
Basic accuracy: 3% (SPC)
Impedance measurement range (SPC)
Measurement time (SPC):
The duration of time between the
triggering of the HP E4916A and the
completion of the measurement was
measured. (The time required to
transfer data is not included. The
DLD measurement was made at 5
points; 10 nW, 100 nW, 1
µ
W, 10
µ
W,
and 100
µ
W.)
Frequency
Q
Fr,CI
Fr,CI+
DLD
MEAS
EQV
TIME
100 MHz
83,000 240 ms
1.02 s 890 ms
SHORT
60 MHz
110,000 190 ms 730 ms 775 ms
SHORT
24 MHz
135,000 146 ms 680 ms 705 ms
SHORT
2 MHz
40,000 280 ms
1.64 s
- -
MEDIUM
2 MHz
- -
- -
- -
2.57 s LONG
Note: Measurement time may vary
depending on the resonance
frequency even if two resona-
tors of the same type are
tested.
Accessories and Other
Equipment
HP E4915A Crystal Impedance
Meter
Inexpensive model (no variable
power function available with the HP
E4916A)
Frequency range: 1 MHz to 180 MHz
Output power: -5dBm fixed
Comes with all HP E4916A measure-
ment capabilities other than DLD,
EM, and LCR measurement functions.
HP 41900A
πππππ
Network Test
Fixture
Meets the
IEC444 and JIS
standards. Ideal
for final and
acceptance
inspections.
Can test
resonators with load capacitance
when used in combination with
option 001.
HP41901A SMD
πππππ
Network Test
Fixture
SMD type
π
network test
fixture that
offers superb
repeatability.
Ideal for final
and acceptance
inspections.
Supports 6 types of QIAJ (Quartz
Crystal Industry Association of
Japan) SMDs.