Specifications - A

51

Table A-2. Supplemental Characteristics (continued)

Parameter

Agilent 66312A

Agilent 6611C -

6614C

Dynamic Measurement
Accuracy

Instantaneous Voltage:

Instantaneous Current:

0.03% + 5 mV

0.6% + 1 mA

1

not applicable

Dynamic Measurement
System

Buffer Length:

Sampling Rate Range:

4096 points

15.6

µ

s

31,200s

not applicable

Measurement Time
(voltage or current)

50 ms average

(includes the default time of 30 ms

2

for acquiring

data, and a 20 ms data processing overhead)

Command Processing
Time

4 ms average

(for output to begin to change

following receipt of digital data)

Remote Sense Capability

Up to 2 V can be dropped across each load lead.

(add 2 mV to the voltage load regulation

specification for each 1 V change in the positive

output lead due to load current change.)

Savable Instrument States
(applies only in SCPI mode)

4 (in locations 0 to 3)

RS-232 Interface
Capabilities

Baud rates:

Data formats:

Language:

300 600 1200 2400 4800 9600

7 bits even or odd parity; 8 bits without parity

SCPI or COMPatibility

3

GPIB Interface
Capabilities

Language:

Interface:

SCPI or COMPatibility

3

AH1, C0, DC1, DT1, E1, L4, PP0, RL1, SH1,

SR1, T6

INH/FLT Characteristics

Maximum ratings:

FLT Terminals:

INH Terminals:

16.5 Vdc between terminals 1 and 2; 3 and 4; and

from terminals 1 or 2 to chassis ground

Low-level output current = 1.25 mA max.

Low-level output voltage = 0.5 V max.

Low-level input voltage = 0.8 V max.

High-level input voltage = 2 V min.

Low-level input current = 1 mA

Pulse width = 100

µ

s min.

Time delay = 4 ms typical

Digital I/O Characteristics

Maximum ratings:

Digital OUT Port 0,1,2

(open collector)

same as INH/FLT Characteristics

Output leakage @ 16V = 0.1 mA (ports 0,1)

= 12.5 mA (port 2)

Output leakage @ 5V = 0.1 mA (ports 0,1)

= 0.25 mA (port 2)

Low-level output sink current @ 0.5 V = 4 mA

Low-level output sink current @ 1 V = 50 mA

1

For full scale current changes with a risetime of 20

µ

s, an additional 0.5% error exists in the first data point in the buffer after the change. The

error percentage increases proportionally with the decrease in risetime.

2

This time may be reduced by changing the default conditions of 2048 data points, however, measurement accuracy will be reduced.

3

COMPatibility language is used to program the Agilent 663xA Series power supplies.