Specifications – Atec Agilent-66312A User Manual

Page 2

background image

49

A

Specifications

Specifications

Table A-1 lists the specifications of the dc source. Specifications are warranted over the ambient
temperature range of 0 to 55

°C. Unless otherwise noted, specifications apply when measured at the rear

terminals after a 30-minute warm-up period.

Table A-1. Performance Specifications

Parameter

Agilent

6611C

Agilent 6612C

Agilent 66312A

Agilent

6613C

Agilent

6614C

Output Ratings

Voltage:

Current:

0 – 8 V

0-5 A

0

− 20 V

0

− 2 A

0 – 50 V

0 –1 A

0 – 100 V

0 – 0.5 A

Programming Accuracy

(@ 25

°C ±5°C)

Voltage: 0.05% +

+Current: 0.05% +

5 mV

2 mA

1

10 mV

1 mA

1

20 mV

0.75 mA

1

50 mV

0.5 mA

1

DC Measurement
Accuracy
(via GPIB or front panel
meters with respect to
actual output @ 25

°C

±5°C)

Voltage: 0.03% +

Low Current range

−20 mA to +20 mA: 0.1% +

High Current range

+20 mA to +rated I: 0.2% +
−20 mA to − rated I: 0.2% +

2 mV

2

2.5

µA

3

0.5mA

4

1.1mA

3 mV

2

2.5

µA

3

0.25 mA

4

0.85 mA

6 mV

2

2.5 uA

3

0.2 mA

4

0.8 mA

12 mV

2

2.5

µA

3

0.1 mA

4

0.7 mA

Ripple and Noise
(in the range of 20 Hz to
20 MHz with outputs
ungrounded or with
either terminal grounded)

Voltage (rms/p-p):

Current (rms):

0.5 mV/3 mV

2 mA

0.5 mV/3 mV

5

1 mA

0.5 mV/4 mV

1 mA

0.6 mV/5 mV

1 mA

Load Regulation

6

(change in output voltage
or current for any load
change within ratings)

Voltage:

Current:

2 mV

1 mA

2 mV

0.5 mA

4 mV

0.5 mA

5 mV

0.5 mA

Line Regulation
(change in output voltage
or current for any line
change within ratings)

Voltage:

Current:

0.5 mV

0.5 mA

0.5 mV

0.5 mA

1 mV

0.25 mA

1 mV

0.25 mA

Transient Response
Time

6

< 100

µs

(for the output voltage to recover to its previous level within 0.1% of the voltage rating of the unit or 20 millivolts following a
change in load current of up to 50% of the output current rating)

1

0.05% + 3.32mA (Agilent 6611C), 1.53mA (Agilent 66312A/12C), 1.01mA (Agilent 6613C), 0.63mA (Agilent 6614C) when programming

between zero and 0.03% of full scale current.

2

Applies for output voltages greater than 10mV (Agilent 6611C/12C), 25mV (Agilent 6613C), and 50mV (Agilent 6614C).

3

This specification may degrade slightly when the unit is subjected to an RF field

≥3 V/meter.

4

For Agilent 66312A: applies in SCPI mode, with current detector set to DC. With current detector set to ACDC, accuracy is 0.2% + four times

the fixed error value. In COMPatibility mode, accuracy is 0.2% + six times the fixed error value.

5

For Agilent 66312A (from 1 MHz to 20 MHz ) = 0.5mV/15mV.

6

Applies at rear terminals with unit set to remote sensing and with sense terminals externally jumpered to their respective output terminals.