2 usb-b interface (usb device), 2 usb-b interface – Xylem ProLab 3000 User Manual
Page 162

Transmitting data
ProLab 3000
162
ba75645e03
12/2012
12.2.2 USB-B interface (USB Device)
Connect the interface to the PC via the supplied Z875 USB cable.
Installation of the USB
driver on the PC
System requirements of the PC for installation of the USB driver:
PC with Pentium processor or higher with at least one free USB con-
nection and CD-ROM drive
Windows 2000, XP.
Note
If the burette control is active, data output to the RS232 interface
(printer) is blocked.
Note
Pin assignment of the socket: See page 179.
1
Insert the supplied installation CD in the CD drive of your PC.
2
Connect the meter to the USB-B input (USB Device) of the PC
via the USB cable.
Windows automatically installs the driver for the meter. Follow
the Windows installation instructions as necessary.
The meter is listed as a virtual COM interface among the con-
nections in the Windows instrument manager.
3
Set up similar transmission data on the PC/printer:
Baud rate: selectable in the range 1200 ... 115200
Handshake: RTS/CTS + Xon/Xoff
Set on the PC only:
– Parity: none
– Data bits: 8
– Stop bits: 1