2 dissolved oxygen – Xylem MultiLab 4010-3 User Manual
Page 90
What to do if...
MultiLab 4010-3
90
ba76143e01
04/2013
No stable measured
value
Obviously incorrect
measured values
13.2 Dissolved oxygen
Error message,
OFL
Error message,
Error
Cause
Remedy
IDS pH sensor:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Adjust temperature if necessary
IDS pH sensor + test sample:
– Conductivity too low
– Use suitable IDS pH sensor
– Temperature too high
– Use suitable IDS pH sensor
– Organic liquids
– Use suitable IDS pH sensor
Cause
Remedy
IDS pH sensor:
– IDS pH sensor unsuitable
– Use suitable IDS sensor
– Temperature difference between
buffer and test sample too great
– Adjust temperature of buffer or
sample solutions
– Measurement procedure not suit-
able
– Follow special procedure
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– Measured value outside the measur-
ing range
– Select a suitable measuring
medium
Cause
Remedy
– Sensor contaminated
– Clean the sensor