3 calibration interval, 2 orp measurement settings, Calibration interval – Xylem MultiLab 4010-3 User Manual

Page 65: Orp measurement settings

background image

MultiLab 4010-3

Settings

ba76143e01

04/2013

65

9.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol.

After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC

FUNCTION

(

SENSOR

QUALITY

CONTROL

), page

40).

After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.

Setting the calibra-

tion interval

The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):

9.2

ORP measurement settings

Settings

The settings are made in the menu for measuring settings of the ORP measure-
ment. To open the settings, display the required measured parameter in the
measured value display and press the key. After completing the set-
tings, switch to the measured value display with .
Default settings are printed in bold.

To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.

1.

Open the menu for measurement settings with .

2.

In the Calibration / Calibration interval menu, set the calibration interval
with <><>.

3.

Confirm the setting with .

4.

Quit the menu with .

Menu item

Possible setting

Explanation

mV resolution

0.1
1

Resolution of the mV display

Reset

-

Resets all sensor settings to the
delivery condition (see section
9.6.1 R

ESETTING

THE

MEASUREMENT

SETTINGS

, page 73).