3 calibration interval, 2 orp measurement settings, Calibration interval – Xylem MultiLab 4010-3 User Manual
Page 65: Orp measurement settings

MultiLab 4010-3
Settings
ba76143e01
04/2013
65
9.1.3 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC
40).
After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.
Setting the calibra-
tion interval
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
9.2
ORP measurement settings
Settings
The settings are made in the menu for measuring settings of the ORP measure-
ment. To open the settings, display the required measured parameter in the
measured value display and press the
tings, switch to the measured value display with
Default settings are printed in bold.
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
1.
Open the menu for measurement settings with
2.
In the Calibration / Calibration interval menu, set the calibration interval
with <><>.
3.
Confirm the setting with
4.
Quit the menu with
Menu item
Possible setting
Explanation
mV resolution
0.1
1
Resolution of the mV display
Reset
-
Resets all sensor settings to the
delivery condition (see section
9.6.1 R