Xylem MultiLab 4010-3 User Manual
Page 56
Conductivity
MultiLab 4010-3
56
ba76143e01
04/2013
You can start the Stability control manually at any time, irrespective of the set-
ting for automatic Stability control (see section 9.5.3 A
, page 72) in the System menu.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
1.
Freeze the measured value with
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD func-
tion with
2.
Using
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured
parameter flashes.
As soon as a measured value meets the stability criteria, it is frozen. The
[HOLD][AR] status indicator is displayed, the progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
ally with
is prematurely terminated, the current measurement data are out-
put to the interface (PC, USB memory device or USB printer) with-
out AutoRead info.
3.
Using
or
Release the frozen measured value again with
The [AR] status display disappears. The display switches back to the
previous indication.
Measured
parameter
Time interval
Stability in the time interval
Conductivity
ϰ
10 seconds
Δ
ϰ: better than
1.0% of measured value
Temperature
15 seconds
Δ
: Better than 0.5 °C