Xylem MultiLab 4010-3 User Manual
Page 30

pH value
MultiLab 4010-3
30
ba76143e01
04/2013
Continuing with two-
point calibration
4.
Thoroughly rinse the sensor with deionized water.
5.
Immerse the sensor in the first buffer solution.
6.
For measurements without temperature sensor
(e.g. when using an IDS adapter
)
:
Enter the temperature of the buffer with <><>.
7.
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
8.
Wait for the end of the measurement with stability control or accept the
calibration value with
The calibration display for the next buffer appears (voltage display).
9.
If necessary, finish the calibration procedure as a single-point calibration
with
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
03.04.2013 08:00
YSI
03.04.2013 08:00
YSI
10.000
10. Thoroughly rinse the sensor with deionized water.
11. Immerse the pH sensor in buffer solution 2.