3 d.o. measurement settings, 1 settings for d.o. measurement, D.o. measurement settings – Xylem MultiLab 4010-3 User Manual
Page 66: Settings for d.o. measurement, Section 9.3.1 s, Ettings, Measurement
Settings
MultiLab 4010-3
66
ba76143e01
04/2013
9.3
D.O. measurement settings
9.3.1 Settings for D.O. measurement
The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the
tings, switch to the measured value display with
Menu item
Possible
setting
Explanation
Calibration /Calibra-
tion record
-
Displays the calibration record of the
last calibration.
Calibration / Cali-
bration data storage
/ Display
-
Displays the last calibration records
(max. 10)
Calibration / Cali-
bration data storage
/ Output to USB
flash drive or printer
-
Outputs the stored calibration data to
the USB-A interface (USB memory
device/USB printer)
Calibration / Cali-
bration data storage
/ Output to RS232/
USB
-
Outputs the stored calibration data to
the USB-B interface (PC)
Calibration /
Calibration data
storage / Number of
cal. points
1
2
1-point calibration
or
2-point calibration
Calibration /Calibra-
tion interval
1 ... 180 ...
999 d
Calibration interval for the D.O. sensor
(in days).
The meter reminds you to calibrate
regularly by the flashing sensor symbol
in the measuring screen.
Calibration /Com-
parison meas.
on
off
Enables to adjust the measured value
with the aid of a comparison measure-
ment, e.g. Winkler titration.
For details, see section 7.2 C
Sal automatic
on
off
Automatic salt content correction for
concentration measurements. The
measured salinity value is taken over
by a connected conductivity sensor.
This menu item is only available if an
IDS conductivity sensor is connected.