beautypg.com

6 reset, 1 resetting the measurement settings, Reset – Xylem MultiLab 4010-3 User Manual

Page 73: Resetting the measurement settings, Section 9.6.1 r, Esetting, Measurement, Settings, Eset, Ting

background image

MultiLab 4010-3

Settings

ba76143e01

04/2013

73

9.6

Reset

You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.

9.6.1 Resetting the measurement settings

pH

The following settings for pH measurements are reset to the default settings
with the Reset function:

The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
key.

ORP

The following settings for ORP measurements are reset to the default settings
with the Reset function:

The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
key.

The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.

Setting

Default settings

Buffer YSI

Calibration interval

7 d

Unit for slope

mV/pH

Measured parameter

pH

pH resolution

0.001

mV resolution

0.1

Asymmetry

0 mV

Slope

-59.2 mV

Man. temperature

25 °C

Single-point calibration

off

Setting

Default settings

mV resolution

0.1

Man. temperature

25 °C