What to do if ...

6 - 1

ba76039e01

01/2012

6

What to do if ...

The sensor

automatically switches

itself on and off

periodically after

commissioning

Mechanical damage to

the sensor

Display of OFL

Display of "----"

Measured value

fluctuating heavily

Measured values too low

Cause

Remedy

– The available power is

sufficient for the initialization of
the sensor, but not for the
operation of the cleaning
system. As soon as this is
switched on, communication
with the controller is
interrupted.

– Install another power supply

module as close as possible
to the VisoTurb

®

700 IQ (SW)

Cause

Remedy

– Return the sensor

Cause

Remedy

– Measuring range exceeded

– See log book

Cause

Remedy

– Measured value invalid

– See log book

Cause

Remedy

– Gas bubbles in the medium are

in front of the sapphire disc

– Check the installation position

of the sensor (see section 3.2
and section 3.3)

– Signal averaging time too short

for low values of turbidity

– Increase signal averaging

time

Cause

Remedy

– Sapphire disc polluted

– Clean sapphire disc (see

section 5.2)