beautypg.com

YSI VisoTurb 700 IQ User Manual

Page 24

background image

Commissioning

VisoTurb

®

700 IQ (SW)

3 - 12

ba76039e01

01/2012

AutoRange
0 ... 0.400 ppm
0 ... 4.00 ppm
0 ... 40.0 ppm
0 ... 400 ppm
0 ... 4000 ppm

Measuring range for the measuring mode
ppm SiO2

(AutoRange = automatic changeover of the
measuring range)

AutoRange
0 ... 0.400 mg/l
0 ... 4.00 mg/l
0 ... 40.0 mg/l
0 ... 400 mg/l
0 ... 4.00 g/l
0 ... 40.0 g/l
0 ... 400 g/l

Measuring range for the measuring mode
g/l TSS

(AutoRange = automatic changeover of the
measuring range)

TSS range

Adjustment values from the reference
measurement for determining the amount of
total suspended solids. Only displayed if the
g/l TSS measuring mode was selected (for
details, see section 4.2.3).

TSS value

Turbidity range

Turbidity value

Signal averaging

1 ... 600 secs

Response time of the signal filter.
Depending on the sample matrix, the
measured values may oscillate more or less
(e.g. due to foreign bodies or air bubbles).
The signal filter reduces the limits of variation
of the measured value. The signal filter is
characterized by the signal averaging time.
This is the time after which 90 % of a signal
change is displayed.

Application offset

-20.00 ... +20.00
(units depend on
measuring mode)

Correction value for compensation of
environmentally-dependent interferences.
The value is added to the measured value
(for details, see section 4.2.2).

Ultrasonic cleaning

On
Off
Pulse

Switch on or off the ultrasound cleaning
function (Pulse = pulse operation).

Save and quit

The system confirms the saving of the
settings and the display switches to the next
higher level.

Setting

Selection/values

Explanation

This manual is related to the following products: