beautypg.com

3 installation examples, Installation examples -5 3.3.1 – YSI VisoTurb 700 IQ User Manual

Page 17

background image

VisoTurb

®

700 IQ (SW)

Commissioning

3 - 5

ba76039e01

01/2012

3.3

Installation examples

As a rule, the VisoTurb

®

700 IQ (SW) will measure interference-free

when the distances and angles etc. specified are observed. However,
interferences at the measuring location (see section 3.2.1) may require
special adaptations of the installation.

3.3.1

Measuring in an open basin or channel (range > 100 FNU)

Example:

Outlet of the

preclarification stage

The turbidity sensor can be immersed in the sample using a pendulum
mounting assembly, e.g. pendulum mounting assembly EH/P 170,
(pay attention to the minimum immersion depth).

Alternatively, the sensor can be suspended on a chain (e.g. with the
EH/F 170 swing mounting assembly and EH/U 170 sensor holder).
Make sure that the sensor cannot bump against any walls or obstacles.

This manual is related to the following products: