Example:
90 ° pipe installation
Fig. 3-6
Turbidity sensor in the pipe (90 °)
The following points must be observed for a right-angled installation in
the pipe (Fig. 3-6):
Rotate the sensor so that the marking on the sensor points in the
direction of the pipe axis
Select a position where the pipe diameter is as large as possible as
the installation location (see section 3.2.5 D
).
Before measuring, check whether an application offset is required in
the case of lower turbidity values (see section 4.2.2).
Marking aid
1
Connect the SACIQ (SW) sensor connection cable to the plug
head connector of the sensor and screw it tight (see section
3.4.1).
2
Attach a marking aid (adhesive strips or similar) in the same
position as the marking on the sensor to the plug head
connector.