The total suspended solids measurement is even more precise, the
better the current status of the sample corresponds to the status at the
time of the user calibration. If there is a fundamental change of the
characteristics of the sample, a new user calibration may be necessary.
Turbidity range
0 ... 0.400 FNU
0 ... 4.00 FNU
0 ... 40.0 FNU
0 ... 400 FNU
0 ... 4000 FNU
Range for the entry of the turbidity value.
Select the smallest possible range in order to
enter the turbidity value in the Turbidity value
field as precisely as possible.
Example:
Reading of the turbidity value= 38.2 FNU
– Smallest possible settings range:
0 ... 40.0 FNU.
– Entry in the Turbidity value field:
38.2 FNU
Turbidity value
Turbidity value as it was determined with the
turbidity sensor. The input precision
depends on the setting in the Turbidity range
field.
Setting
Selection/values
Explanation
11
Make the sensor settings with
d
and confirm each of them
with
g
.
12
Select the Save and quit menu item with
d
and confirm with
g
. The new settings are stored in the sensor. The turbidity
sensor is calibrated for the measurement of total suspended
solids.