beautypg.com

YSI IQ S ENSOR N ET MIQ/CHV PLUS User Manual

Page 24

background image

Settings

MIQ/CHV PLUS

4 - 8

ba76036e01

01/2012

Note
The reference time and all further cleaning times relate to the date and
time of the system clock. How to set the system clock is described in
the system operating manual.

Example

Fig. 4-5

Cleaning cycle

Setting

Result

Reference time (h):
Reference time (min):
Interval unit:
Cleaning interval:

12
0
Hours (h)
8 h

Reference time: 12:00 hours
This specifies the following
start times:
04:00, 12:00 and 20:00 hours

1

Reference time
Start of a cleaning cycle (t1)
Start of the specified Cleaning duration (t1a)

2

End of the specified Cleaning duration (t1a)
Start of the specified Adjustment time (t1b)

3

End of the specified Adjustment time (t1b)
End of the cleaning cycle (t1)

4

Reference time ± Cleaning interval (t2)
Start of a cleaning cycle

t1

Cleaning cycle =
Cleaning duration (t1a) Adjustment time (t1b)
Linked outputs are frozen.

t2

Cleaning interval

relay
condition

open

closed

00:00

Time

t1

t1

t1

4

4

t2

t2

t2

t2

1

04:00

12:00

24:00

20:00

1

2

t1a t1b