YSI IQ S ENSOR N ET MIQ/CHV PLUS User Manual
Page 23

MIQ/CHV PLUS
Settings
4 - 7
ba76036e01
01/2012
Settings
* With short cleaning intervals, the adjustable values for the Cleaning 
duration and Adjustment time are limited. The following values apply:
Note
With this, the cleaning times are fixed. They only change when the 
reference time is changed.
Setting
Selection/Values
Explanation
Reference time (h)
0 ... 23 h
Time at which a 
cleaning cycle is 
started. Further 
cleaning cycles will 
be performed at the 
times specified by the 
cleaning interval.
Reference time 
(min)
0 ... 60 min
Interval unit
1 .. 7 d
1 .. 24 h
5 .. 60 min
Selection of range 
and unit for the 
Cleaning interval.
Cleaning interval
1/2/3/4/5/6/7 d
or:
1/2/3/4/6/8/12/24 h
or:
5/10/15/20/30/60 min
Repeat interval for 
the cleaning function: 
Time between the 
start time of a 
cleaning cycle and 
the start time of the 
next cleaning cycle*.
Cleaning duration
0 ... 300 s
Duration of the 
cleaning
Adjustment time
0 ... 900 s
Time extension to 
allow the sensor to 
adjust to the test 
sample after 
cleaning.
Cleaning interval
Cleaning duration
Adjustment time
≤
10 min
max. 60 s
max. 120 s
≤
20 min
max. 180 s
max. 300 s