YSI DIQ/S 182 XT-4 System 182 XT-4 User Manual
Page 9
System 182 XT-4
Overview
1 - 3
ba76026e01
01/2012
Relay and current
outputs
The relay and current outputs can be linked with sensors. Linked
outputs can be used to monitor sensors and for the output of measured
values.
A relay output can be programmed as:
a monitoring relay
a limit monitor
a proportional output of measured values (pulse width or frequency
output)
the control unit of a sensor cleaning system driven by compressed
air.
A current output can be programmed as:
an analog output
a PID controller
For quick orientation, the states of all relay and current outputs are
clearly indicated on the display.
Cleaning system driven
by compressed air
The DIQ/S 182 XT-4 is prepared for the time-controlled sensor
cleaning function driven by compressed air. The cleaning process is
controlled from the operating unit. The operating unit provides the
supply voltage and control relay for the compressed air valve in the
DIQ/CHV valve module. This enables simple setup and uncomplicated
wiring.
Alternatively, the MIQ/CHV PLUS valve module can be installed in the
system. It combines the relay, valve and valve power supply in one MIQ
enclosure. Thus, no additional wiring is required, which makes
installation easier especially if there is a greater distance between the
DIQ/S 182 XT-4 and sensor.
Further system options
In case of increased energy demand, an additional power supply unit
can be installed for the power supply if necessary.