5 setting tables for current outputs, 1 functions and settings, 2 recorder – YSI DIQ/S 182 XT-4 System 182 XT-4 User Manual
Page 118: Setting tables for current outputs -26 6.5.1, Functions and settings -26
Outputs
System 182 XT-4
6 - 26
ba76026e01
01/2012
6.5
Setting tables for current outputs
6.5.1
Functions and settings
To set a current output, its function has to be selected first. Then the
relevant setting table is displayed:
Function of current
outputs and relevant
setting tables
6.5.2
Recorder
Function
The measured values of the linked sensor at the current output are set
up as current intensity in the Recorder application. The output of
measured values is laid down in the Recorder type, Start value and End
value settings.
Settings
Function
Setting table
No function
The current output is not used.
Recorder
PID controller
Fixed current value
Setting
Selection/Values
Explanation
Recorder type
0 to 20 mA or
4 to 20 mA
Start value
Within the
measuring range
(sensor-
dependent)
Minimum spacing:
≤ 5 % of the measuring
range
(sensor-dependent)
End value
Measured variable
Main variable
Adjoining variable
Main variable designates
the actual measured
parameter of the sensor
(e.g. pH, oxygen, etc.).
Adjoining variable
designates an additional
measured parameter
(e.g. temperature).
Attenuation
0 ... 40 mA/s
Speed of change of the
output current (mA/s) in
the case of erratic
changes of the input
signal.