YSI DIQ/S 182 XT-4 System 182 XT-4 User Manual
Page 13

System 182 XT-4
Safety instructions
2 - 3
ba76026e01
01/2012
Safe operation
If safe operation is no longer possible, the System 182 XT-4 must be
taken out of operation and secured against inadvertent operation.
Safe operation is no longer possible if components:
have been damaged in transport
have been stored under adverse conditions for a prolonged period
of time
are visibly damaged
no longer operate as described in this manual.
If you are in any doubt, contact the supplier of your System 182 XT-4.
Obligations of the
operator
The operator of the System 182 XT-4 must ensure that the regulations
and guidelines listed below are followed when dealing with dangerous
substances:
EEC guidelines relating to safety at work
National laws relating to safety at work
Accident prevention regulations
Safety datasheets of the chemicals manufacturers.
Warning
A circuit (except for the power supply connections) that is
connected to the System 182 XT-4 must not feed any voltages or
currents that are not allowed. The circuit must comply with all the
standards of a circuit with limitations (Limited Circuit or Limited
Power) as well as for safety extra low voltage (SELV) at all times.
These include the following limiting value specifications:
AC voltage: max. 30 V effective / 42.4 V peak
DC voltage: max. 60 V
Current limit: max. 8 A
Power output limitation: max. 150 VA