19. Appendix D - Remote Command Language
Remote commands and data output are available only when the instrument is in the Remote
mode.
Command Format:
Prefix:
The prefix consists of the two character sequence
Suffix:
The suffix consists of the two character sequence
Command Types and Parameters:
Command types are arranged into two groups, immediate and automatic.
The Display command controls the display of remote messages.
Command type parameter description
D
0
string
echoes the string to the first line of the display
D
1
string
echoes the string to the second line of the display
D
2
none
Clear display lines 0 and 1
D
3
0
Set display update flag line 0
D
3
1
Set display update flag line 1
The ouTput mode command controls how data flow is controlled. A bar code scan in T1 or T2
mode will cause a line of output, just as if the T command had been received.
Command type parameter description
T
0
none
Continuous data output
T
1
none
One line of data output in response
T
2
none
One line after stable reading