Press [SETUP] to change the SOUR parameters. The following screen shows the default
parameters.
Sample / Total 1 / [ 1]
SOUR
Min. Time
[ 1] min.
Max. Time
[ 15] min.
[*]SOUR@20
Min. Beginning DO
[ 5.00] mg/L SEND:
off
Min. Ending DO
[ 2.00] mg/L
Solids Weight
[ 1.000] g/L
UP
DOWN
DIGIT
NEXT
Use the [UP], [DOWN], [DIGIT] and [NEXT] soft-keys to change the following parameters as
necessary:
Sample / Total:
Enter the ratio of sample volume to total volume. For example, if
you are diluting 1 to 10 (1 part sample plus 9 parts dilution water)
then enter the number 10. This would be a 10-fold dilution.
Min. Time:
Enter the minimum time (in minutes).
Max. Time:
Enter the maximum time (in minutes). When the maximum time is
reached, the test will end.
Min. Beginning DO: Enter the minimum level of DO allowed at the start of the test. If the
DO falls below this level, the test will not start.
Min. Ending DO:
Enter the minimum level of DO allowed during the test. If the DO
level falls to this value, the test will end.
Solids Weight:
Enter the Total Solids or Volatile Suspended Solids concentration of
the sample in g/L. The maximum acceptable value is 31.999 g/L.
SOUR@20:
This applies the calculation for temperature correction to 20°C
according to the Farrell and Bhide equation as follows:
SOUR
20
= SOUR
T
x Θ
(20-T)
Where:
Θ = 1.05 above 20°C
1.07 below 20°C
This calculation is only valid for temperature ranges from 10° - 30°C
SEND: on/off
This sets the RS232 status at the beginning of the test.
After the parameters have been set, press [ENTER] to save them and return to the SOUR menu.