12. Troubleshooting
Error Messages
The instrument performs a Power On Self Test each time it is turned on. The following error
codes are provided to facilitate troubleshooting.
Code/Problem Error
Possible
Cause
Correction
Blank Display
Dead/low batteries
Replace batteries or
connect power supply
5011
adapter
improperly connected
Unplug 5011 from
power supply connector
E1: Check Probe
Temperature under
range
Connector improperly
installed
Check probe
connection
Intermittent
connection
in cable or plug
Repair or replace
Temperature
over
range
Electrical leakage in
connector or cable
Repair or replace
Faulty
temperature
sensor
Repair or replace
E2: Check
Membrane
High Background
current (DO)
Insufficient warm-up
time
See Calibration
Incorrect probe zeroing
procedure
See Zero Calibration
Probe needs servicing
See probe instructions
Probe malfunction
Repair or replace
E3: Check
Membrane
DO Low current
Insufficient electrolyte
See probe instructions
Contaminated
electrodes or fouled
membrane
See probe instructions
Membrane too thick
Try another membrane
High resistance in
probe connection
Repair or replace