beautypg.com

YSI 5000 User Manual

Page 39

background image

36

If you want to save these values press the [STORE] soft-key. A record number will then show
up under the REVIEW menu. If the record slot contains a set of data, an indicator saying FULL
will appear beside the test number. If the record slot does not contain a set of data, the indicator
will say EMPTY.

REVIEW

62.9 mg /L /hr

8.5 mg /L

REC:1 EMPTY

11:04 AM

09/21/98

NEXT REC.

MODIFY

SAVE

SEND

Press the [NEXT REC.] soft-key to change the record number to find an empty slot, or pick a slot
to overwrite (you cannot delete records, you must overwright). Press the [SAVE] soft-key to
save the current test data into that slot. Once you have saved the data you can press the
[MODIFY] soft-key to change some parameters of the test.

MODIFY

50.0 mg /L /hr

5.6 mg /L

REC:1 FULL

11:04 PM

09/21/98

!

"

SETTINGS

Press the ! or " soft-keys to move the start time reference point to an optimal place on the
curve. The reference point is represented by a vertical line. Only data to the right of the
reference point will be used to calculate the OUR value. Moving this reference point allows you
to start the test at anytime and trim off unwanted data after the test has run. The DO reading at
the reference point is shown at the top. The reference point cannot move closer than the
minimum time to the end of the test.

The [SETTINGS] soft-key allows you to modify or input values for the Dilution factor. Changes
to the Dilution factor are saved if you save the record again.

0

mg
/L

870

OUR

DO (at Reference

Test time

Time Reference

0

mg
/L

870

Record Number

FULL or EMPTY
indicator

This manual is related to the following products: