YSI 5000 User Manual
Page 39
36
If you want to save these values press the [STORE] soft-key. A record number will then show
up under the REVIEW menu. If the record slot contains a set of data, an indicator saying FULL
will appear beside the test number. If the record slot does not contain a set of data, the indicator
will say EMPTY.
REVIEW
62.9 mg /L /hr
8.5 mg /L
REC:1 EMPTY
11:04 AM
09/21/98
NEXT REC.
MODIFY
SAVE
SEND
Press the [NEXT REC.] soft-key to change the record number to find an empty slot, or pick a slot
to overwrite (you cannot delete records, you must overwright). Press the [SAVE] soft-key to
save the current test data into that slot. Once you have saved the data you can press the
[MODIFY] soft-key to change some parameters of the test.
MODIFY
50.0 mg /L /hr
5.6 mg /L
REC:1 FULL
11:04 PM
09/21/98
!
"
SETTINGS
Press the ! or " soft-keys to move the start time reference point to an optimal place on the
curve. The reference point is represented by a vertical line. Only data to the right of the
reference point will be used to calculate the OUR value. Moving this reference point allows you
to start the test at anytime and trim off unwanted data after the test has run. The DO reading at
the reference point is shown at the top. The reference point cannot move closer than the
minimum time to the end of the test.
The [SETTINGS] soft-key allows you to modify or input values for the Dilution factor. Changes
to the Dilution factor are saved if you save the record again.
0
mg
/L
870
OUR
DO (at Reference
Test time
Time Reference
0
mg
/L
870
Record Number
FULL or EMPTY
indicator