Xylem handylab OX 12 User Manual
Page 9
handylab Ox 12
Overview
9
ba75369e04
01/2011
Accuracy
(± 1 digit)
Correction
functions
Analog output
Automatic switchover when a recorder is connected.
Concentration
mg/l
± 0.5 % of the measured value
at an ambient temperature of 5 °C ... 30 °C
%
saturation
± 0.5 % of the measured value
when measuring in the range of
± 10 K around the calibration temperature
Partial pressure
mbar
± 0.5 % of the measured value
at an ambient temperature of 5 °C ... 30 °C
Temperature
compensation
< 2 % at 0 ... 40 °C
T [°C]
± 0.1
Salinity correction
0 ... 70.0 SAL
Air pressure correction
Automatic through installation of pressure
sensor in the range 500 ... 1100 mbar
Concentration
Voltage
Resolution
0 ... 19.99 mg/l
0 ... 1999 mV
0.01 mg/l per 1 mV
0 ... 90.0 mg/l
0 ... 900 mV
0.1 mg/l per 1 mV
Saturation
Voltage
Resolution
0 ... 199.9
0 ... 1999 mV
0.1 % per 1 mV
0 ... 600 %
0 ... 600 mV
1 % per 1 mV
D. O. partial pressure
Voltage
Resolution
0 ... 199.9 mbar
0 ... 1999 mV
0.1 mbar per 1 mV
0 ... 1250 mbar
0 ... 1250 mV
1 mbar per 1 mV
Accuracy
± 0.5 % of the displayed value ± 0.1 (% sat-
uration)
or
± 0.5 % of the displayed value ± 0.01 mg/l