Xylem handylab OX 12 User Manual

Page 9

background image

handylab Ox 12

Overview

9

ba75369e04

01/2011

Accuracy
(± 1 digit)

Correction

functions

Analog output

Automatic switchover when a recorder is connected.

Concentration
mg/l

± 0.5 % of the measured value
at an ambient temperature of 5 °C ... 30 °C

%
saturation

± 0.5 % of the measured value
when measuring in the range of
± 10 K around the calibration temperature

Partial pressure
mbar

± 0.5 % of the measured value
at an ambient temperature of 5 °C ... 30 °C

Temperature
compensation

< 2 % at 0 ... 40 °C

T [°C]

± 0.1

Salinity correction

0 ... 70.0 SAL

Air pressure correction

Automatic through installation of pressure
sensor in the range 500 ... 1100 mbar

Concentration

Voltage

Resolution

0 ... 19.99 mg/l

0 ... 1999 mV

0.01 mg/l per 1 mV

0 ... 90.0 mg/l

0 ... 900 mV

0.1 mg/l per 1 mV

Saturation

Voltage

Resolution

0 ... 199.9

0 ... 1999 mV

0.1 % per 1 mV

0 ... 600 %

0 ... 600 mV

1 % per 1 mV

D. O. partial pressure

Voltage

Resolution

0 ... 199.9 mbar

0 ... 1999 mV

0.1 mbar per 1 mV

0 ... 1250 mbar

0 ... 1250 mV

1 mbar per 1 mV

Accuracy

± 0.5 % of the displayed value ± 0.1 (% sat-
uration)
or
± 0.5 % of the displayed value ± 0.01 mg/l