handylab Ox 12

What to do if...

49

ba75369e04

01/2011

6

What to do if...

LoBat display

Instrument does

not react to key-

stroke

Error message

0FL

Error message

E3

Cause

Remedy

– Batteries almost empty

– Replace batteries

(see section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition

undefined or EMC load
unallowed

– Processor reset:

Press the and
keys at the same
time and release them again.
The software version is
displayed.

Cause

Remedy

Display range exceeded

D. O. probe:

– Not connected

– Connect the probe

– Cable broken

– Replace probe

– Depleted

– Replace probe

– Short-circuit between gold

and lead electrode

– Clean probe and replace it if

necessary

Cause

Remedy

Invalid calibration

D. O. probe:

– Electrolyte solution depleted

– Regenerate probe

– Membrane contaminated

– Clean membrane

– Electrode system poisoned

– Regenerate probe

– Worn out

– Replace probe

– broken

– Replace probe