1 calibrating with the air calibration vessel, Calibrating with the air calibration vessel, A l c – Xylem handylab OX 12 User Manual
Page 25
handylab Ox 12
Operation
25
ba75369e04
01/2011
4.3.1
Calibrating with the air calibration vessel
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet).
1
Connect the D. O. probe to the measuring instrument.
2
Keep an air calibration vessel ready for use.
3
Put the D. O. probe into the air calibration vessel.
4
Press the
2
CAL appears.
5
Press the
6
As soon as a stable value is reached, the AR display indicator stops
flashing. This indicates that the calibration is complete; the sensor
symbol shows the relative slope that was determined and the sensor
evaluation (see page 24).