handylab LF 12

What to do if...

53

ba75367e04

01/2011

6

What to do if...

LoBat display

Instrument does

not react to key-

stroke

Error message

OFL

Error message

E3

Display

to

Probe symbol

flashes

Cause

Remedy

– Batteries almost empty

– Replace batteries

(see section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition

undefined or EMC load
unallowed

– Processor reset:

Press the and
keys at the same
time and release them again.
The software version is
displayed.

Cause

Remedy

The measured value lies outside
the measuring range

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and

replace it if necessary

– Unsuitable calibration

solution

– Check calibration solutions

Cause

Remedy

– Time-out of the interface

– Check the instrument that is

connected

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system