handylab LF 12
Operation
43
ba75367e04
01/2011
4.5.2
Recorder (analog output)
You can transmit data to a recorder via the analog output. Connect the ana-
log output to the recorder via the Z 394 interface cable.
The data output automatically switches to Recorder output.
Socket
assignment
Note
The analog output is activated automatically in the cable by connecting 2 and
3.
The signal range of the analog output depends on the measured variable and
the measuring range:
Conductivity/
Resistance
Salinity
TDS
1 free
2 Plug coding
3 Ground
4 Analog output
1
2
3
4
RS 232
REC
(internal resistance < 5 Ohm)
Measuring range
Voltage
Resolution
0.000 ... 1.999 µS/cm
0 ... 1999 mV
1 mV
0.00 ... 19.99 µS/cm
0 ... 1999 mV
1 mV
0.0 ... 199.9 µS/cm
0 ... 1999 mV
1 mV
0 ... 1999 µS/cm
0 ... 1999 mV
1 mV
0.00 ... 19.99 mS/cm
0 ... 1999 mV
1 mV
0.0 ... 199.9 mS/cm
0 ... 1999 mV
1 mV
0 ... 500 mS/cm
0 ... 500 mV
1 mV
Measuring range
Voltage
Resolution
0 ... 70.0
0 ... 700 mV
1 mV
Measuring range
Voltage
Resolution
0 ... 1999 mg/l
0 ... 1999 mV
1 mV