Operation

handylab LF 12

24

ba75367e04

01/2011

Note
If the error message

E3 appears, refer to chapter 6 W

HAT

TO

DO

IF

...

4

The displayed value is the current, calibrated cell constant. You
can:
– accept this setting for measuring with or

– continue with step 5 and start a new calibration.

5

Immerse the measuring cell in the control standard solution,
0.01 mol/KCl.

6

Press the

key.

– If no temperature sensor is connected, enter the current

temperature of the solution with <V> <W>
and confirm with .

– If a temperature sensor is connected, the AR measurement

to determine the cell constant starts.

The AR display indicator flashes until a stable signal is
reached. The cell constant determined is displayed. The mea-
suring instrument automatically stores the cell constant.