Operation

23

ba75367e04

01/2011

4.3.1

Determining the cell constant
(calibrating in the control standard)

Note
This method of automatically determining the cell constant by calibration with
the 0.01 mol/l KCL standard solution can only be used for measuring cells
with cell constants in the range 0.450 ... 0.500 cm

-1

or 0.800 ... 1.200 cm

-1

.

This is how you can determine the cell constant:

Display

Cell constant [cm

-1

]

0.450 ... 0.500 cm

-1

0.800 ... 1.200 cm

-1

E3
Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

outside the ranges
0.450 ... 0.500 cm

-1

or
0.800 ... 1.200 cm

-1

1

Press the key until CELL appears on the display.

2

Press the key.

3

Press the key repeatedly, until the calibrated cell con-
stant appears on the display: CAL appears on the display.