Operation
handylab LF 12
34
ba75367e04
01/2011
4.4.2
Saving automatically
The save interval (Int 1) determines the chronological interval between auto-
matic save processes.
After the fixed interval has expired, the current data record is transmitted to
the storage and to the interface.
Setting the save
interval
The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min, 10 min,
15 min, 30 min, 60 min):
1
Press the
2
Press the
3
Set the required interval between the saving procedures with
<V> <W>.
4
Confirm with
The number of free memory locations appears on the display.
5
As soon as all of the 800 storage locations are occupied,
AutoStore is terminated (Int 1 = OFF).
If there are not enough storage locations available for your
measurements:
– Output and backup the data storage (see page 36) and
– clear the memory (see page 40).