Display
Sensoface
Gives you indications about the status of the
sensor, analysis of zero-point/slope, setting time,
frequency of calibration. Status indication
(friendly, neutral, sad)
Display
LCD STN 7-segment display with 3 lines and
icons
Status indicators
Battery power level, logger
Notices
Hourglass
Keypad
[on/off], [cal], [meas], [set], [], [ ], [STO], [RCL],
[clock]
Diagnostics functions
Sensor data
(Memosens only)
Manufacturer, sensor type, serial number,
operating time
Calibration data
Calibration date, zero, slope
Device self-test
Automatic memory test (FLASH, EEPROM, RAM)
Device data
Device type, software version, hardware version
Data retention
Parameters, calibration data > 10 years
Data transfer
1x Micro USB-B for data transfer to PC
Never use the USB port within the hazardous location.
Data logger
With up to 5000 memory locations
Calibration data logger
MemoLog (Memosens only)
Recording: manual, interval- or event-controlled
Up to 100 Memosens calibration protocols recordable
- Records displayable
- Directly transferable via Memosuite (USB): manufacturer, sensor type, serial no.,
zero-point, slope, calibration date
Communication
USB 2.0
Profi le
HID, installation without driver
Use
Data exchange and confi guration via HandyLab
Pilot software
Nominal operating
conditions
Ambient temperature
–10 … +55 °C
Transport/ Storage temperature –25 … +70 °C
Relative humidity
0 … 95 %, short-term condensing allowed
Power
Power supply
Batteries 4x AA alkaline or 4x NiMH
(rechargeable)
Operating time
approx 1000 h (Alkaline)
Housing
Material
PA12 GF30 + TPE
Protection
IP66/67 with pressure compensation
Dimensions
approx. (132 x 156 x 30) mm
Weight
approx. 500 g
Licensing - Compliance - Equipment safety
EMC
EN 61326-1 (General Requirements)
Emitted interference
Class B (residential area)
Immunity to interference
Industry
EN 61326-2-3 (Particular Requirements for Transmitters)
Specifi cations