Data Logger

Operating Modes of the Data Logger ( Logger Type)

Manual logging when logger is activated ( SHOT)

In this mode, a measured value is recorded when the STO key is pressed.

Measurement
Logger activated

The measured value is saved to the address of the last recorded value + 1

Manual logging when logger is deactivated

Measurement
Logger deactivated

Measured value is maintained
Proposed address blinks
(address of the last recorded
value + 1)

If desired: Select start address
using

.

Measured value is saved to the desired address (e.g. for overwriting an
incorrect measurement).

Interval (INT)

In this mode, the measured values are cyclically recorded.

Example: Logger type “INT”

Time

= entry

pH