beautypg.com

Safety guidelines – Xylem HandyLab 750(EX) User Manual

Page 37

background image

37

37

Safety guidelines

(HandyLab 750 EX)

Warning!

• Only open the battery compartment of the HandyLab 750, 750

EX outside the hazardous location.

• Never try to open the device. If a repair should be required,

return the device to our factory.

• Never use the USB port within the hazardous location.

Warning!

When using the HandyLab 750 EX (device variant for applica-
tions in hazardous locations) in a hazardous location, only the
battery types listed below may be used. The batteries must be
from the same manufacturer and of identical type and capaci-
ty. Never use new and used batteries together.

Batteries for Application in Hazardous Locations

Batteries (4x each)

Temp. class

Ambient temperature range

Duracell MN1500

T4

-10 °C ≤ Ta ≤ +40 °C

Energizer E91

T3

-10 °C ≤ Ta ≤ +50 °C

Power One 4106

T3

-10 °C ≤ Ta ≤ +50 °C

Panasonic Pro Power LR6 T3

-10 °C ≤ Ta ≤ +50 °C

Precautions for application in hazardous locations

In hazardous areas, only the HandyLab 750 EX may be used. The following
safety guidelines must be observed:

This manual is related to the following products: