HandyLab 200

What to do if...

43

ba77063e01

03/2014

6

What to do if...

Error message

OFL, UFL

Error message,

Error

Sensor symbol flashes

Display

Meter does not react to

keystroke

You want to know which

software

version is in the meter

Cause

Remedy

– Measured value outside the

measuring range

– Use suitable measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Calibration solution not suitable

– Check the calibration solu-

tions

– Measuring cell not connected

– Connect the measuring cell

Cause

Remedy

– Cleaning interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Batteries almost empty

– Replace the batteries (see

section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the and
key simultane-
ously

Cause

Remedy

– E.g., a question by the service

department

– Switch on the meter.

Open the menu,

_

>[Menu] / Storage &

config / System / Service
information
. The instrument
data are displayed.