HandyLab 200

Operation

29

ba77063e01

03/2014

4.4.4

Settings for conductivity measuring cells

Overview

The following settings are possible for conductivity measuring cells:

 Calibration record (display)

 Calibration interval

 Measuring cell/cell constant

 Reference temperature

 Temperature compensation

 TDS factor

 Unit of the temperature

 Automatic Stability control

Settings

The settings are made in the Measurement menu for the measured
parameter, conductivity. To open the settings, display the required
parameter in the measured value display and press the /[menu] or
key. After completing the settings, switch to the measured
value display with .

Menu item

Possible set-
ting

Explanation

Calibration /
Calibration record

-

Displays the calibration
record of the last calibra-
tion.

Calibration /
Calibration interval

1 ... 999 d

Calibration interval for the
measuring cell (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring screen.