Xylem HandyLab 200 User Manual
Page 30
![background image](/manuals/310606/30/background.png)
Operation
HandyLab 200
30
ba77063e01
03/2014
Measurement /
Measuring cell /
Type
Measuring cell used
Cal
Measuring cells the cell
constant of which is deter-
mined by calibration in the
KCL control standard solu-
tion.
Calibration ranges:
0.450 to 0.500 cm
-1
and
0.800 to 0.880 cm
-1
The currently valid cell con-
stant is displayed in the sta-
tus line.
LF313T
Measuring cell LF313T,
nominal cell constant
0.100 cm
-1
.
The cell constant can be
adjusted in the range from
0.090 to 0.110 cm
-1
.
LF213T
Measuring cell LF213T,
nominal cell constant
0.010 cm
-1
.
The cell constant is perma-
nently set.
man
Any measuring cells with
freely adjustable cell con-
stants in the range from
0.250 to 25.000 cm
-1
.
Measurement /
Measuring cell /
Cell const. man
0.250 to
25.000 cm
-1
Display and setting option
of the cell constant of any
measuring cells (man).
Measurement /
Measuring cell /
Cell const. LF313T
0.090 to
0.110 cm
-1
Display and setting options
for the cell constant of the
measuring cell LF313T.
Measurement /
Temp. comp. (TC) /
Method
nLF
Lin
Off
Procedure for temperature
compensation (see section
4.4.3).
This setting is only avail-
able for the measured
parameters,
ϰ
and
ρ
.
Menu item
Possible set-
ting
Explanation