3 temperature compensation, Temperature compensation – Xylem HandyLab 200 User Manual
Page 27
HandyLab 200
Operation
27
ba77063e01
03/2014
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.4.3
Temperature compensation
The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
Tr20 or Tr25.
You can select one of the following temperature compensation meth-
ods:
2
Using
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized.
Note
You can prematurely terminate the Stability control function manually
with
turely terminated, the current measurement data are output to the
interface without the AutoRead info.
3
Using
control.
or
Release the frozen measured value again with
The [AR] status display disappears. The display switches back
to the previous indication.
Measured
parameter
Time interval
Stability in the time interval
Conductivity
ϰ
10 seconds
Δ
: better than 1.0% of mea-
sured value
Temperature
15 seconds
Δ
: better than 0.5 °C