beautypg.com

Xylem HandyLab 200 User Manual

Page 28

background image

Operation

HandyLab 200

28

ba77063e01

03/2014

Nonlinear temperature compensation (nLF) according to EN 27

888

Linear temperature compensation (lin) with adjustable coeffi-

cients of 0.000 ... 3.000 %

/

K

 No temperature compensation (off)

Note
The reference temperature and temperature compensation are set in
the Measurement menu for the parameter, conductivity (see section
4.4.4
).

Application tips

Select the following temperature compensations given in the table
according to the respective test sample:

Test sample

Temperature compensation

Display
indica-
tor

Natural water
(ground water, surface
water, drinking water)

nLF
according to EN 27 888

nLF

Ultrapure water

nLF
according to EN 27 888

nLF

Other aqueous solu-
tions

lin Set linear
temperature coefficient
0.001 ... 3.000 %/K

lin

Salinity (seawater)

Automatic nLF according to
IOT (International Oceano-
graphic Tables)

Sal, nLF