4 Specifications *
Accuracy:
< 1% of FS range under constant conditions
Analysis:
0-1 ppm, 0-10, 0-100, 0-1000 ppm FS ranges;
auto-ranging or manually lock on single range
Application:
Analyze oxygen concentrations from 10 ppb to 1000 ppm
in inert, hydrocarbon, helium, hydrogen and gas streams
Approvals:
CE, Intrinsic Safety (pending)
Area Classification: Meets recognized intrinsic safety standards for use in Class
1, Division 1, Group C, D hazardous areas (see Options)
Calibration:
Certified gas of O
2
balance N
2
approximating 80% of
analysis range or one range above analysis range
Compensation:
Temperature and barometric pressure
Connections:
1/8" compression tube fittings
Controls:
Water resistant keypad; menu driven range selection,
calibration and system functions
Display:
Graphical LCD 2.75 x 1.375; resolution .001 ppb; displays
real time ambient temperature and pressure
Enclosure:
Painted aluminum NEMA 4X, 8.6x9x3", 12 lbs.
Flow Sensitivity:
None between 1-3 SCFH, 1 SCFH recommended
LED Indicators:
LOW BATT (72 hr. warning); CHARGE mode
Linearity:
> .995 over all ranges
Pressure:
Inlet - regulate to 5-30 psig; vent - atmospheric
Power:
Rechargeable battery, 60 day duty cycle (pump 1 day)
Recovery Time:
60 seconds in air to < 1 ppm in < 20 minutes on N
2
purge
Response Time:
90% of final FS reading in 10 seconds
Sample System:
Flow control, bypass and isolation valves; flow indicator
Sensitivity:
< 0.5% of FS range
Sensor Model:
GPR-12-2000MS
Sensor Life:
24 months at 25ºC and 1 atm; average O
2
< 100 ppm
Signal Output:
0-1V
Temp. Range:
5º to 45ºC
Warranty:
12 months analyzer; 12 months sensor
Wetted Parts:
Stainless steel
Optional Equipment
Integral sampling pump - general purpose or intrinsically safe designs
Carrying case with custom foam insert
Sample conditioning accessories - contact factory
* Specifications subject to change without notice
GPR-1200 MS
Portable ppm O
2
Analyzer
Advanced Sensor Technology
Accuracy < 1% FS Range
Sensitivity < 0.5% FS Range
Fast Recovery to < 1 ppm
24 Month Expected Life
No Maintenance
Intrinsically Safe Design
4 Standard Analysis Ranges
Auto or Manual Ranging
SS Bypass Sample System
Water Resistant Enclosure
ISO 9001:2008 Approved QA System
Integral bypass sample system increases productivity.