beautypg.com

Maintenance – Analytical Industries GPR-1200 MS ATEX Portable PPB Oxygen Analyzer User Manual

Page 5

background image

5

Maintenance

Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the operator to

service.
Only trained personnel with the authorization of their supervisor should conduct maintenance.
Oxygen Sensor: DO NOT open the sensor. The sensor contains a corrosive liquid electrolyte that could be

harmful if touched or ingested, refer to the Material Safety Data Sheet contained in the Owner’s Manual appendix.

Avoid contact with any liquid or crystal type powder in or around the sensor or sensor housing, as either could be a

form of electrolyte. Leaking sensors should be disposed of in accordance with local regulations.
Troubleshooting: Consult the guidelines in Section 8 for advice on the common operating errors before
concluding that your analyzer is faulty.
Do not attempt to service the analyzer beyond those means described in this Owner’s Manual. Do not attempt to

make repairs by yourself as this will void the warranty as per Section 10 and may result in electrical shock, injury or

damage. All other servicing should be referred to qualified service personnel.
Cleaning: The analyzer should be cleaned only as recommended by the manufacturer. Wipe off dust and dirt from

the outside of the unit with a soft damp cloth then dry immediately. Do not use solvents or chemicals.
Nonuse Periods: If the analyzer is equipped with a range switch advance the switch to the OFF position and

disconnect the power when the analyzer is left unused for a long period of time.

This manual is related to the following products: