2 quality control certification – Analytical Industries GPR-2500 A Oxygen Analyzer User Manual
Page 4

4
2 Quality Control Certification
Date:
Customer: Order No.:
Pass
Model:
GPR-2500A Oxygen Transmitter
Sensor:
( ) GPR-11-32-4 Oxygen Sensor
( ) XLT-11-24-4 oxygen Sensor
Serial Nos.:
Transmitter ______________________ Sensor ______________________
Accessories:
Owner’s Manual
ENCL-1071-KIT Y-Fitting Packing Fiber, Sealing Cement, Stick Stirrer
Configuration:
A-1151-E-A2 PCB Assembly Main Processing
A-1153-A2 PCB Assembly Alarms/Power Connection
Software rev:
Ranges: 0-1%, 0-5%, 0-10%, 0-25%
Power: 12-36V DC two wire loop power
Barometric pressure and temperature compensation
NEMA 4X rated wall mount enclosure
Test: Pass
Pass
Set default zero
Set default span @ 40uA
Alarm delay
Zero calibration
Alarm bypass
Span Calibration
Alarm configurations; ALARM 1, ALARM 2
Analog signal output 4-20mA full scale
Alarm function; ALARM 1, ALARM 2
Calibrates with adequate span adjustment within 10-50% FS
Alarm relays; ALARM 1, ALARM 2
Baseline drift on zero gas <
± 2% FS over 24 hour period
Alarm system fail, dry contact
Noise level <
± 1.0% FS
Overall inspection for physical defects
Options:
Notes: