beautypg.com

Start-up – Analytical Industries GPR-2500 A Oxygen Analyzer User Manual

Page 11

background image



11









Example: As illustrated by Graph A any error, play in the multi-turn span pot or the temperature compensation

circuit, during a span adjustment at 20.9% (air) of full scale range would be multiplied by a factor of 4.78

(100/20.9) if used for measurements of 95-100% oxygen concentrations. Conversely, an error during a span

adjustment at 100% of full scale range is reduced proportionately for measurements of lower oxygen

concentrations.
Recommendation: Calibrating with a span gas approximating 80% of the full scale range one or two ranges
higher than the full scale range of interest is recommended for 'optimum calibration accuracy'. Always calibrate at

the same temperature and pressure of the sample gas stream.

Start-up

The GPR-2500A Oxygen Transmitter has been tested, calibrated at the factory prior to shipment with the sensor

installed and is fully operational from the shipping container. Allow the transmitters to stabilize for 30 minutes and

then recalibrate the device as instructed below.
Installation Considerations:
The GPR-2500A consists of a two circuit boards, sensor housing and sample 1/8” sample inlet and vent connections
housed in a NEMA 4X rated enclosure. The complete transmitter package measures 9Hx4Wx3.5”D and is suitable

for mounting on any vertical flat surface.
For optimum accuracy zero and calibrate a transmitter after it has been allowed to stabilize, typically 24-36 hours

after installation. Assuming the initial zero is performed according to the procedure described herein, the analyzer

should not require zeroing again until the either the sensor is replaced or a change is made to the sample system

or gas lines. Following the initial zero and calibration, the analyzer should not require span calibration again for up
to 3 months under “normal” application conditions as described in the published specifications.

Note: As described below, zeroing the transmitter is recommended only for measurements below 1% and not
practical for measurement ranges above 1%. The low end sensitivity (zero capability) has been verified at the

factory; however, no ZERO OFFSET adjustment has been made. A factory adjustment would be meaningless

because of the difference in sample systems and leakage factors between the factory set-up and the actual

application conditions.

¾ Assemble the necessary hardware for mounting the transmitter and optional components - such as coalescing

or particulate filters and pumps, 1/8” metal or plastic tubing for interconnecting the transmitter and optional

components.

¾ Temperature: The sample must be sufficiently cooled before it enters the transmitter and any optional

components. A coiled 10 foot length of ¼” stainless steel tubing is sufficient for cooling sample gases as high

as 1,800ºF to ambient.