Advanced instruments inc – Analytical Industries GPR-1600 MS Series PPB Oxygen Analyzer User Manual
Page 43

Advanced Instruments Inc.
 
 
 
43
Symptom
Possible Cause
Recommended Action
 
Erratic O
2
reading
 or 
No O
2
reading
 
 
 
 
 
 
 
 
Change in sample pressure 
 
 
 
Dirty electrical contacts in upper section of 
sensor housing 
 
Corroded solder joints on sensor PCB from 
corrosive sample or electrolyte leakage from 
sensor 
Corroded spring loaded contact in upper 
section of sensor housing from liquid in 
sample or electrolyte leakage from sensor 
 
 
 
 
Liquid covering sensing area 
 
Improper sensor selection 
 
Presence of interference gases 
Unauthorized maintenance 
Sensor nearing end of life 
 
 
Sensors without PCB use mV setting. 
Calibrate the analyzer (calibrate at pressure and 
temperature of sample) 
 
Clean contacts with alcohol (minimize exposure time 
of MS sensor to ambient air to extent possible) 
 
Replace sensor and return sensor to the factory for 
warranty determination 
 
Upper section of sensor housing: Clean contacts 
with alcohol, flow sample or zero gas for 2-3 hours 
to flush sample system and sensor housing 
Sensor: Replace if leaking and return it to the 
factory for warranty determination 
 
Wipe with alcohol and lint free towel or flow sample 
or zero gas for 2-3 hours to flush 
Replace GPR/PSR sensor with XLT sensor when CO
2
or acid gases are present. Consult factory. 
 
Replace sensor and install scrubber 
Consult factory. 
Replace sensor 
 
 
Erratic O
2
reading
 or 
Negative O
2
reading
 or 
No O
2
reading
accompanied by 
electrolyte leakage 
 
 
Pressurizing the sensor by flowing gas to the 
sensor with the vent restricted or SHUT OFF 
valve closed and suddenly removing the 
restriction draws a vacuum on the sensor 
 or 
partially opening the valves upstream of the 
analyzer when using a pump downstream of 
the analyzer to draw sample from a process at 
atmospheric pressure or a slight vacuum. 
Placing a vacuum on the sensor in excess 4” 
of water column is strongly discouraged. 
 
A premature adjustment of the ZERO OFFSET 
potentiometer is a common problem 
 
 
Zero the analyzer. If not successful replace the 
sensor 
 
 
 
Avoid drawing a vacuum on the sensor, a 
pressurized sensor may not leak but still produce 
negative readings. 
 
 
 
 
 
From MAIN MENU select DEFAULT ZERO 
