beautypg.com

Advanced instruments inc, Menu navigation, Main menu – Analytical Industries GPR-1600 MS Series PPB Oxygen Analyzer User Manual

Page 23: Range selection

background image

Advanced Instruments Inc.



23

START-UP TEST


ELECTRONICS – PASS
TEMP SENSOR – PASS
BAROMETRIC SENSOR – PASS

REV. 1.61

Note: The analyzer display defaults to the sampling mode when 30 seconds elapses without user interface.
Note: At installation expect the range to default to 25% range, thereafter, 100 ppm range but only if properly isolated.

3.3

PPM

AUTO SAMPLING

10 PPM RANGE

24.5 C

LO1 2 PPM 10 PPM HI2


Menu Navigation

The four (4) pushbuttons located on the front of the analyzer operate the micro-processor:
1. blue ENTER (select)
2. yellow UP ARROW
3. yellow DOWN ARROW
4. green MENU (escape)






Main Menu

Access the MAIN MENU by pressing the MENU key:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE
CALIBRATION
CONFIG ALARMS
BYPASS ALARMS


Range Selection

The GPR-1600MS analyzer is equipped with four (4) standard measuring ranges (see specification) and provides users with a
choice of sampling modes. By accessing the MAIN MENU, users may select either the AUTO SAMPLING (ranging) or MANUAL
SAMPLING (to lock on a single range) mode.