Advanced instruments inc – Analytical Industries GPR-1600 MS Series PPB Oxygen Analyzer User Manual
Page 31

Advanced Instruments Inc.
 
 
 
31
CALIBRATION 
SPAN CALIBRATE 
ZERO CALIBRATE 
DEFAULT SPAN
DEFAULT ZERO 
OUTPUT SPAN 
OUTPUT ZERO 
 
11. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN. 
12. Press the ENTER key to select the highlighted menu option. 
13. The following display (below left) appears and after 3 seconds the system returns to the SAMPLING mode: 
 
 
 
FACTORY
DEFAULTS
SET
 
 
 
 
 
3.3
PPM
AUTO SAMPLING
10 PPM RANGE
24.5 C 
LO1 2 PPM 10 PPM HI2 
 
14. Repeat steps 1 through 4 above. 
 
CALIBRATION 
SPAN CALIBRATE 
ZERO CALIBRATE
DEFAULT SPAN 
DEFAULT ZERO 
OUTPUT SPAN 
OUTPUT ZERO
15. Advance the reverse shade cursor using the ARROW keys to highlight ZERO CALIBRATE. 
16. Press the ENTER key to select the highlighted menu option. 
17. The following displays appear: 
 
0.000 PPM
 
ZERO 
CALIBRATION 
ENTER TO CALIBRATE 
MENU TO ABORT 
 
 
 
18. Press the ENTER key to calibrate or MENU key to abort and return to SAMPLING mode. 
19. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or 
reading has stabilized within 50% of the full scale low range.
20. Both the Zero Calibrate and Span Calibrate functions result in the following displays: 
 
 
 
 
PASSED
CALIBRATION
OR
 
 
FAILED
CALIBRATION
 
 
 