Advanced instrtuments inc, 5 operation - appendix a – Analytical Industries GPR-3100 Series Oxygen Purity Analyzer User Manual

Page 41

background image

Advanced Instrtuments Inc.

5 Operation - Appendix A

* MAIN MENU Sample
Sample
Span
Zero
Alarm
System

Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48

99.55 %

MAIN MENU Sample

ZERO Sample

* MAIN MENU Sample

Sample

* Factory Default (3)

Sample

(3) ZERO menu, Factory Default option -

Span

Calibrate

Span

Purpose is to reset the microprocessor as the first step in

* Zero

Zero

troubleshooting perceived abnormalities. The zero value

Alarm

Alarm

(which may be compromised) is reset using an average

System

System

sensor output value to temporarily assess the analyzer's
performance. Zero calibration is required upon completion.

Auto Range 90 to 100 %

Auto Range 90 to 100 %

Auto Range 90 to 100 %

75F 101 Kpa 02/10/2004 17:32:48

75F 101 Kpa 02/10/2004 17:32:48

75F 101 Kpa 02/10/2004 17:32:48

Move cursor (*) to Zero using DOWN ARROW key

Move cursor (*) to Factory Default using DOWN ARROW key

Press ENTER to select

Press ENTER to select
Activates (1) and returns to MAIN MENU

Note: Returns to MAIN MENU if no entry within 30 seconds

ZERO Sample

* MAIN MENU Sample

Factory Default

Sample

* Calibrate

Span
Zero
Alarm
System

Auto Range 90 to 100 %

6.50 %

Auto Range 0 to 100 %

75F 101 Kpa 02/10/2004 17:32:48

Enter to Cal, ESC to Abort

O2 Content High, Please Wait

75F 101 Kpa 02/10/2004 17:32:48

Move cursor (*) to Calibrate using DOWN ARROW key

Press ENTER to select Calibrate

(4) Above display remains until ppm oxygen value reaches

Press ENTER to select

6% of analyzer's lowest range. Adjustment is made and
returns to MAIN MENU

* MAIN MENU Sample
Sample
Span
Zero
Alarm
System

Auto Range 90 to 100 %

Enter to Cal, ESC to Abort

75F 101 Kpa 02/10/2004 17:32:48

Press ESC to select Abort

Timesout and returns to MAIN MENU

99.55 %

99.55 %

00.05 %

ZERO

CALIBRATION

IN

PROGRESS

ZERO

CALIBRATION

IN

PROGRESS

ZERO

CALIBRATION

IN

PROGRESS

CALIBRATION

FAILED

99.55 %

5_ops_3100_menu 11/22/2005 Page 3 of 7