Advanced instrtuments inc, 5 operation - appendix a – Analytical Industries GPR-3100 Series Oxygen Purity Analyzer User Manual
Page 41

Advanced Instrtuments Inc.
5 Operation - Appendix A
* MAIN MENU Sample
Sample
Span
Zero
Alarm
System
Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48
99.55 %
MAIN MENU Sample
ZERO Sample
* MAIN MENU Sample
Sample
* Factory Default (3)
Sample
(3) ZERO menu, Factory Default option -
Span
Calibrate
Span
Purpose is to reset the microprocessor as the first step in
* Zero
Zero
troubleshooting perceived abnormalities. The zero value
Alarm
Alarm
(which may be compromised) is reset using an average
System
System
sensor output value to temporarily assess the analyzer's
performance. Zero calibration is required upon completion.
Auto Range 90 to 100 %
Auto Range 90 to 100 %
Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48
75F 101 Kpa 02/10/2004 17:32:48
75F 101 Kpa 02/10/2004 17:32:48
Move cursor (*) to Zero using DOWN ARROW key
Move cursor (*) to Factory Default using DOWN ARROW key
Press ENTER to select
Press ENTER to select
Activates (1) and returns to MAIN MENU
Note: Returns to MAIN MENU if no entry within 30 seconds
ZERO Sample
* MAIN MENU Sample
Factory Default
Sample
* Calibrate
Span
Zero
Alarm
System
Auto Range 90 to 100 %
6.50 %
Auto Range 0 to 100 %
75F 101 Kpa 02/10/2004 17:32:48
Enter to Cal, ESC to Abort
O2 Content High, Please Wait
75F 101 Kpa 02/10/2004 17:32:48
Move cursor (*) to Calibrate using DOWN ARROW key
Press ENTER to select Calibrate
(4) Above display remains until ppm oxygen value reaches
Press ENTER to select
6% of analyzer's lowest range. Adjustment is made and
returns to MAIN MENU
* MAIN MENU Sample
Sample
Span
Zero
Alarm
System
Auto Range 90 to 100 %
Enter to Cal, ESC to Abort
75F 101 Kpa 02/10/2004 17:32:48
Press ESC to select Abort
Timesout and returns to MAIN MENU
99.55 %
99.55 %
00.05 %
ZERO
CALIBRATION
IN
PROGRESS
ZERO
CALIBRATION
IN
PROGRESS
ZERO
CALIBRATION
IN
PROGRESS
CALIBRATION
FAILED
99.55 %
5_ops_3100_menu 11/22/2005 Page 3 of 7